When blood comes into contact with dialyzer membranes, it can induce pathophysiological reactions. The biocompatible membrane in Diacap® CE protects your patients by drastically reducing this reaction potential.
Using a special unique polymer coupling technology, a polymer chain coating is applied to the blood-side of the dialyzer membrane surface. These polymer chains consist of hydrophilic polyethylene glycol (PEG) groups and hydrophobic alkyl groups. By bending and swaying on the membrane surface, these chains create a diffusive layer which physically prevents blood components from coming into direct contact with the underlying membrane surface. Blood-membrane interaction is markedly reduced.
The Diacap® CE membrane meets the sophisticated demands of a gentle and high efficient low flux dialysis.
Benefits:
- Suppresses complement activation
- Reduces thrombogenicity
- Suppresses platelet activation
- Lowers protein adsorption
Technical Data
| In vitro performance |
Diacap® CE 1.3 |
Diacap® CE 1.5 |
Diacap® CE 2.0 |
| Ultrafiltration coefficient (ml/h . mmHg) |
7.8 |
8.8 |
12.0 |
Clearances: QB 200 ml/min
Urea
Creatinine
Phosphate
Vitamin B12 |
184
166
142
66 |
188
172
150
74 |
195
190
163
89 |
Clearances: QB 300 ml/min
Urea
Creatinine
Phosphate
Vitamin B12 |
237
203
166
70 |
246
214
178
79 |
266
229
198
97 |
Clearances: QB 400 ml/min
Urea
Creatinine
Phosphate
Vitamin B12 |
269
225
179
72 |
282
238
193
82 |
311
258
218
101 |
| Surface (m2) |
1.3 |
1.5 |
2.0 |
| Wall Thickness / Internal diameter (μm) |
10/180 |
| Priming volume (ml) bloodside |
78 |
94 |
123 |
| Membrane material |
Alkyl ether polymer grafted
cellulose |
| Housing |
Styrene copolymer |
| Potting Compound |
Polyurethane |
| Sterilisation |
Gamma-ray |
| Units per box |
12 |
| Art-No |
7208000 |
7208001 |
7208002 |
|
|